Development of a Differential Mobility Analyzer for a Titan Lander, 15-R8816

Principal Investigators
Kristin Favela
Adam Hawley
Amy McCleney
Gregory Miller
Jordan Nielson
Edward Patrick
Inclusive Dates 
01/01/18 - Current

Background

The purpose of this project was to simulate, design, fabricate, and test a differential mobility analyzer (DMA) suitable for a Titan lander. Since the discovery of Titan’s atmosphere in 1944, findings from Voyager 1 and 2 and most recently Cassini-Huygens have pointed to a unique atmospheric chemistry involving nitrogen (N2) and methane (CH4). This unique chemistry leads to tholins, thought to be responsible for the brown/orange haze in the atmosphere. The Ion Neutral Mass Spectrometer (INMS) of Cassini-Huygens measured C3H8, C4H2, HCN, HC3N, C6H6, and C2N2. The Cassini Plasma Spectrometer (CAPS), Ion Beam Spectrometer (IBS), and Electron Spectrometer (ELS) revealed what is suspected to be heavy hydrocarbon and nitrile species with the IBS detecting positive ions up to 350 Daltons (Da) and the ELS measuring negative ions up to 13,800 Da. Interest in tholins has sparked many laboratory simulations on creating and analyzing tholins to try and explain Titan’s chemistry. Our motivation for building a DMA is that a DMA analyzes ions based on their size-to-charge, or mobility, thereby providing information on molecular and aerosol sizes of tholins in the atmosphere of Titan.

Approach

The differential mobility analyzer (DMA) is a parallel plate design (Figure 1). The two parallel plates are called the inlet and outlet plates with the inlet plate containing the inlet slit for ions to enter the DMA and the outlet plate having an outlet slit at a fixed horizontal distance (x-dimension) from the inlet slit. In this design, a voltage difference is applied between the two plates providing an electric field to accelerate ions (z-dimension) between the plates while a perpendicular flow of gas pushes the ions along the horizontal axis of the DMA (z-dimension). The mobility of the ion (ion size) determines its collision frequency with the perpendicular flow of gas and how fast it can move in the z-dimension from the applied electric field. Ions of high mobility, considered of small size and low cross section, are able to move a greater distance between the plates compared to larger sized ions of lower mobility at a fixed electric field. For a fixed perpendicular gas flow rate, the electric field between the two plates can be scanned so ions of all mobility can be collected after the exit slit. A DMA plot would show ion current, or ion signal, on the y-axis versus the DMA voltage on the x-axis. Thus, it is a scanning technique where the DMA voltage scan reveals information on the size of ions passing through the device. Coupling the DMA to a mass spectrometer would provide orthogonal separation methods and yield information of ion size and mass-to-charge. This would aid in the understanding of the structure and chemistry of Titan tholins.

cross sectional views of the DMA design with a front-on view (left) and isometric view (right)

Figure 1: Two cross sectional views of the DMA design with a front-on view (at left) and isometric view (at right).

Accomplishments

We have successfully simulated, designed, and fabricated the prototype differential mobility analyzer (DMA). Initial two-dimensional boundary layer thickness calculations were performed to determine optimal DMA contraction design (Görtler parameter) to provide uniform flow. A computational fluid dynamics (CFD) software program was also used to simulate the resolution and sensitivity of the DMA design. A design was selected that could analyze particles in the size range of 1-50 µm at a uniform gas flow rate and reasonable Reynolds number. The prototype DMA is shown in Figure 2. An electrospray ionization (ESI) source has been developed to introduce ions to the DMA. The ESI source has been successfully tested (Figure 3) and provides ion currents up to tens of nanoamperes. Experiments are currently underway to show proof-of-concept operation through the entire DMA device.

 
DMA prototype

Figure 2: Photograph of the DMA prototype.

As part of the experimental testing, we plan to analyze tholins in the DMA to demonstrate the proof-of-concept for an instrument suitable for Titan. We have successfully produced tholins as a part of this program as confirmed by mass spectrometry analysis and scanning electron micrograph images. A more in-depth experimental endeavor in tholin production, which started from this project, was performed in the project, “A Study on Tholin Production as an Analogue for Titan’s Atmospheric Chemistry” (15-R8819).

Taylor cone from the ESI needle

Figure 3: Photograph of the Taylor cone from the ESI needle.